Health and Happiness : Our Fundamental Rights®

Fully Automated Processor
Fully Automated Processor

Product Description

Robotic hand. Pipetting probe, 1ul stepping (1000ul volumes).Total 25 racks, arbitrary rack can be defined as sample or reagent on the software. 4 units 96 well microplates (pipetting,incubator,reading). 1 unit independent washer &reader( two 8 nozzle manifolds).

 

 




Technical Specifications:
 
Sample pipetting unit Sample rack    24 sample tubes/rack, at most 25 racks
Primary sample tubes    11.5-13mm
Pipetting    8ul~1000ul, (10ul) CV<5%
Time to dispense    15 minutes (typical)^100ul of 96 samples to plate 
Dispensing system    1 aspirating-dispensing probe with X-Y-Z movement;
   dilution system with a   syringe of 1000ul, continuous
   flow washing station.

 
Reagent pipetting unit Reagent racks    2 reagent tanks(60ml)/ rack, at most 25 racks
QC racks    24 positions/QC rack, at most 25 racks
Pipetting    8ul~1000ul,1ul stepping,
pipetting precision    (100ul )CV≤0.5%
Time to dispense
 
   2 minutes(typical)^100ul of 96 reagents to
   plate from reagent bottle


 
Washing unit Programmable    Yes(options & parameters)
Wash containers    4 wash buffers at 2.0L, with level-sensing
Waste containers    10L with waste full sensor
Washing system    two 8 nozzle manifolds, one to dispense
   and the other one to aspirate;
Washing residual
 
   <2μl

 
Reading unit Reading system    8 independent photometric channels with mono
   and bichromatic reading capability
Photometric range    0.000 ~3.000 OD
Spectral range    400-700nm
Optical filters    2 precision filters (450and 630nm); 2 more on request
   (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm)
Reading inaccuracy
 
   inaccuracy±1% from 0.000 to 1.500 OD, ±2%
   from 1.500 to 3.000 OD


 
Incubation unit Incubators    4 independent units
Programmable    Yes
Temperature range    RT to 45 °C 
Temperature accuracy    ±0.5℃
Temperature staility
 
   ±0.5℃

 
Software unit Software system    Windows XP or Windows 7 or above, LIS system
   available,realize two-way data communications
Function application    work list set up, patient reports and result filing;
   calibrationcurve, storage for each carried out test
Data reduction
 
   cut-off (qualitative)
 
Physical Specification Power Supply    AC220V+ 10%,50/60Hz,450W 
Instrument Dimension    110*72*70cm (L*W*H)
Instrument Weight    100kg
Package Dimension    138*91*98cm(L*W*H)
Total Weight    150kg