Auto Biochemistry Analyzer-Emerald | ||
Technical Parameters |
||
Overall Performance |
Type | Discrete, Fully automatic |
Throughput | 800 Tests per hour | |
Analysis Method | End-point, Fixed-time, Rate, Turbidimetry | |
Programming Items | Programmable for any test items | |
Certificate |
CE, FDA, ISO9001, ISO14001 , ISO13485 |
|
Sample unit |
Sample Position | 180 samples, double disk design, including 60 samples for calibration/quality dribbling refrigerated |
Sample Tube | Original blood collection tube, micro tube, plastic tube, etc. | |
Sample Type | Serum, plasma, urine, etc. | |
Sample Volume | 2ul~70ul(0.1ul stepping) | |
Sample Probe | Having liquid level induction function, ensuring the accuracy of adding sample |
|
Probe Cleaning |
Automatic washing both interior and exterior, cross contamination rate<0.1% |
|
Reagent unit |
Reagent Position | 200 reagent positions, dual-tray design |
Wash Solution | Washing solution can be installed in any position, including needle, stir bar washing solution and water bath disinfectant |
|
Reagent Volume | 20-350ul, 0.5ul increment | |
Temperature Control | Continuous 2~8°C cooling within 24 hours | |
Reagent Probe | having liquid level induction function, ensuring the accuracy of adding reagent. Inventory checking |
|
Probe Cleaning | Automatic washing both interior and exterior | |
Analysis Items |
Analyzing maximally 100 items at the same time |
|
Reaction unit |
Reaction Volume | 150-550ul reaction temperature 37°C |
Temperature System | Water bath, Real-time temperature display | |
Cuvette | 160 positions with permanent hard glass cuvettes (optical path length 5mm) |
|
Mixer | Independent stir bar, nanotechnology, highly polished, automatic frequency conversion |
|
Washing | Automatic water cleaning and drying system | |
System adjustment |
System can record the blank value and get more accurate result by deducting blank base value of each cuvette |
|
Optical System |
Light Source | Halogen lamp, lamp can be auto steep and wake up |
Receiver | Holographic flat field concave grating diode array detection | |
Wavelength Range | 16 wavelengths from 340 to 800nm | |
Linear Range | -0.3000~3.3000 Abs | |
Siray light |
< 0.1% or Abs> 3.0 |
|
Calibration and QC |
Calibration Method | Linear calibration: K factor, two-and multiple linear Nonlinear Calibration: spline,polynomial, logarithm, etc. |
Quality Control |
Auto QC,cooled ac positions |
|
Software |
STAT Function | Emergency sample can be placed in the sample tray at any time |
Software | Windows 7/8/10 or above, LIS system available. | |
Online | Full bi-directional communication possible | |
Data Processing |
With reaction curve detection and data display function in the whole process |
|
Others |
Water consumption | Deionized water, <45 Llhour |
Power Supply | 110/220V±10%, 60/50Hz | |
Package Size(W*D*H) | 1720*1200*1430mm | |
Gross Weight | 600kg |